Section 410IAC26-17-4. Maintenance of equipment  


Latest version.
  •    All patient care equipment must be in good working order and regularly serviced and maintained as follows:

    (1) All patient care equipment must be on a documented maintenance schedule of appropriate frequency in accordance with one (1) of the following:

    (A) Acceptable standards of practice.

    (B) The manufacturer's recommended maintenance schedule.

    (2) There must be evidence of preventive maintenance on all patient care equipment.

    (3) Appropriate records must be:

    (A) kept pertaining to:

    (i) equipment maintenance;

    (ii) repairs; and

    (iii) electrical current leakage checks; and

    (B) analyzed at least triennially.

    (4) Defibrillators must be discharged at least in accordance with manufacturers' recommendations, and a discharge log with initialed entries must be maintained.

    (Indiana State Department of Health; 410 IAC 26-17-4; filed May 11, 2006, 9:36 a.m.: 29 IR 3371; readopted filed Jul 12, 2012, 12:08 p.m.: 20120808-IR-410120196RFA)