Section 410IAC26-17-4. Maintenance of equipment
-
All patient care equipment must be in good working order and regularly serviced and maintained as follows:
(1) All patient care equipment must be on a documented maintenance schedule of appropriate frequency in accordance with one (1) of the following:
(A) Acceptable standards of practice.
(B) The manufacturer's recommended maintenance schedule.
(2) There must be evidence of preventive maintenance on all patient care equipment.
(3) Appropriate records must be:
(A) kept pertaining to:
(i) equipment maintenance;
(ii) repairs; and
(iii) electrical current leakage checks; and
(B) analyzed at least triennially.
(4) Defibrillators must be discharged at least in accordance with manufacturers' recommendations, and a discharge log with initialed entries must be maintained.