Section 410IAC5-9-9. Operating and emergency procedures


Latest version.
  •  

                                    (a) Particle accelerators, when not in operation, shall be secured to prevent unauthorized use.

                                    (b) The safety interlock system shall not be used to turn off the accelerator beam except in an emergency.

                                    (c) All safety and warning devices, including interlocks, shall be checked for proper operation at intervals not to exceed three months. Results of such tests shall be maintained at the accelerator for inspection by the board.

                                    (d) Electrical circuit diagrams of the accelerator and the associated safety interlock systems shall be kept current and maintained for inspection by the board and shall be available to the operator at each accelerator facility.

                                    (e) If, for any reason, it is necessary to intentionally bypass a safety interlock or interlocks, such action shall be:

    (1) Authorized by the radiation safety committee and/or radiation safety officer;

    (2) Recorded in a permanent log and a notice posted at the accelerator control console; and

    (3) Terminated as soon as possible.

                                    (f) A copy of the current operating and the emergency procedures shall be maintained at the accelerator control panel. (Indiana State Department of Health; Rule HRH-2, PT I, Sec I.9; filed May 26, 1978, 3:30 pm: 1 IR 226; filed Feb 29, 1984, 10:10 am: 7 IR 975; readopted filed Jul 11, 2001, 2:23 p.m.: 24 IR 4234; readopted filed May 22, 2007, 1:44 p.m.: 20070613-IR-410070141RFA; readopted filed Sep 11, 2013, 3:19 p.m.: 20131009-IR-410130346RFA)